Podatność CVE-2018-8868


Publikacja: 2018-07-02   Modyfikacja: 2018-07-03

Opis:
Medtronic MyCareLink Patient Monitor, 24950 MyCareLink Monitor, all versions, and 24952 MyCareLink Monitor, all versions, contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can apply the other vulnerabilities within this advisory to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.

Typ:

CWE-noinfo

CVSS2 => (AV:L/AC:M/Au:N/C:C/I:C/A:C)

Ogólna skala CVSS
Znaczenie
Łatwość wykorzystania
6.9/10
10/10
3.4/10
Wymagany dostęp
Złożoność ataku
Autoryzacja
Lokalny
Średnia
Nie wymagana
Wpływ na poufność
Wpływ na integralność
Wpływ na dostępność
Pełny
Pełny
Pełny
Affected software
Medtronic -> 24950 mycarelink monitor firmware 
Medtronic -> 24952 mycarelink monitor firmware 

 Referencje:
https://ics-cert.us-cert.gov/advisories/ICSMA-18-179-01