Home
Bugtraq
Full List
Only Bugs
Only Tricks
Only Exploits
Only Dorks
Only CVE
Only CWE
Fake Notes
Ranking
CVEMAP
Full List
Show Vendors
Show Products
CWE Dictionary
Check CVE Id
Check CWE Id
Search
Bugtraq
CVEMAP
By author
CVE Id
CWE Id
By vendors
By products
RSS
Bugtraq
CVEMAP
CVE Products
Bugs
Exploits
Dorks
More
cIFrex
Facebook
Twitter
Donate
About
Submit
Vulnerabilities for
'Plum a+3 infusion system firmware'
2019-03-25
CVE-2015-3956
CWE-345
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
CVE-2015-3954
CWE-285
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
CVE-2015-3953
CWE-798
Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
CVE-2015-3952
CWE-200
Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
>>>
Vendor:
Pifzer
3
Products
Symbiq infusion system firmware
Plum a+3 infusion system firmware
Plum a+ infusion system firmware
Copyright
2024
, cxsecurity.com
Back to Top