Index
Bugtraq
Pełna lista
Błędy
Sztuczki
Exploity
Dorks list
Tylko z CVE
Tylko z CWE
Bogus
Ranking
CVEMAP
Świeża lista CVE
Producenci
Produkty
Słownik CWE
Sprawdź nr. CVE
Sprawdź nr. CWE
Szukaj
W Bugtraq
W bazie CVE
Po autorze
Po nr. CVE
Po nr. CWE
Po producencie
Po produkcie
RSS
Bugtraq
CVEMAP
CVE Produkty
Tylko Błędy
Tylko Exploity
Tylko Dorks
Więcej
cIFrex
Facebook
Twitter
Donate
O bazie
Lang
Polish
English
Submit
Podatności dla
'Plum a+ infusion system firmware'
2019-03-25
CVE-2015-3956
CWE-345
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
CVE-2015-3954
CWE-285
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
CVE-2015-3953
CWE-798
Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
CVE-2015-3952
CWE-200
Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
>>>
Vendor:
Pifzer
3
Produkty
Symbiq infusion system firmware
Plum a+3 infusion system firmware
Plum a+ infusion system firmware
Copyright
2024
, cxsecurity.com
Back to Top