Vulnerability CVE-2014-5432


Published: 2019-03-26

Description:
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes to the WBM, as well as issue commands to access account credentials and shared keys. Baxter asserts that this vulnerability only allows access to features and functionality on the WBM and that the SIGMA Spectrum infusion pump cannot be controlled from the WBM. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.

Type:

CWE-287

(Improper Authentication)

CVSS2 => (AV:N/AC:L/Au:N/C:P/I:P/A:P)

CVSS Base Score
Impact Subscore
Exploitability Subscore
7.5/10
6.4/10
10/10
Exploit range
Attack complexity
Authentication
Remote
Low
No required
Confidentiality impact
Integrity impact
Availability impact
Partial
Partial
Partial
Affected software
Baxter -> Sigma spectrum infusion system firmware 

 References:
https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01

Copyright 2024, cxsecurity.com

 