Vulnerability CVE-2014-5433


Published: 2019-03-26

Description:
An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16, which may allow an attacker to gain access the host network. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.

Type:

CWE-255

(Credentials Management)

CVSS2 => (AV:N/AC:L/Au:N/C:P/I:P/A:P)

CVSS Base Score
Impact Subscore
Exploitability Subscore
7.5/10
6.4/10
10/10
Exploit range
Attack complexity
Authentication
Remote
Low
No required
Confidentiality impact
Integrity impact
Availability impact
Partial
Partial
Partial
Affected software
Baxter -> Sigma spectrum infusion system firmware 

 References:
https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01

Copyright 2024, cxsecurity.com

 

Back to Top