Vulnerability CVE-2014-7233


Published: 2015-08-04

Description:
GE Healthcare Precision THUNIS-800+ has a default password of (1) 1973 for the factory default System Utilities menu, (2) TH8740 for installation using TH8740_122_Setup.exe, (3) hrml for "Setup and Activation" using DSASetup, and (4) an empty string for Shutter Configuration, which has unspecified impact and attack vectors. NOTE: since these passwords appear to be used to access functionality during installation, this issue might not cross privilege boundaries and might not be a vulnerability.

Type:

CWE-255

(Credentials Management)

CVSS2 => (AV:N/AC:L/Au:N/C:C/I:C/A:C)

CVSS Base Score
Impact Subscore
Exploitability Subscore
10/10
10/10
10/10
Exploit range
Attack complexity
Authentication
Remote
Low
No required
Confidentiality impact
Integrity impact
Availability impact
Complete
Complete
Complete
Affected software
Gehealthcare -> Precision thunis-800+ 

 References:
http://apps.gehealthcare.com/servlet/ClientServlet/5458232-1EN+r4.pdf?REQ=RAA&DIRECTION=5458232-1EN&FILENAME=5458232-1EN%2Br4.pdf&FILEREV=4&DOCREV_ORG=4
http://www.forbes.com/sites/thomasbrewster/2015/07/10/vulnerable-breasts/
https://ics-cert.us-cert.gov/advisories/ICSMA-18-037-02
https://twitter.com/digitalbond/status/619250429751222277

Copyright 2024, cxsecurity.com

 