Vulnerability CVE-2017-12720


Published: 2018-02-15

Description:
An Improper Access Control issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The FTP server on the pump does not require authentication if the pump is configured to allow FTP connections.

Type:

CWE-306

(Missing Authentication for Critical Function)

CVSS2 => (AV:N/AC:M/Au:N/C:P/I:P/A:P)

CVSS Base Score
Impact Subscore
Exploitability Subscore
6.8/10
6.4/10
8.6/10
Exploit range
Attack complexity
Authentication
Remote
Medium
No required
Confidentiality impact
Integrity impact
Availability impact
Partial
Partial
Partial
Affected software
Smiths-medical -> Medfusion 4000 wireless syringe infusion pump 

 References:
http://www.securityfocus.com/bid/100665
https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A