Vulnerability CVE-2018-10596
Published: 2018-07-02 Modified: 2018-07-03
Description: |
Medtronic 2090 CareLink Programmer all versions The affected product uses a virtual private network connection to securely download updates. The product does not verify it is still connected to this virtual private network before downloading updates. An attacker with local network access to the programmer could influence these communications. |
Type:
CWE-200 (Information Exposure)
CVSS2 => (AV:A/AC:L/Au:S/C:P/I:P/A:P)
CVSS Base Score |
Impact Subscore |
Exploitability Subscore |
5.2/10 |
6.4/10 |
5.1/10 |
Exploit range |
Attack complexity |
Authentication |
Adjacent network |
Low |
Single time |
Confidentiality impact |
Integrity impact |
Availability impact |
Partial |
Partial |
Partial |
References: |
https://ics-cert.us-cert.gov/advisories/ICSMA-18-058-01
|
|
|