Vulnerability CVE-2018-10596


Published: 2018-07-02   Modified: 2018-07-03

Description:
Medtronic 2090 CareLink Programmer all versions The affected product uses a virtual private network connection to securely download updates. The product does not verify it is still connected to this virtual private network before downloading updates. An attacker with local network access to the programmer could influence these communications.

Type:

CWE-200

(Information Exposure)

CVSS2 => (AV:A/AC:L/Au:S/C:P/I:P/A:P)

CVSS Base Score
Impact Subscore
Exploitability Subscore
5.2/10
6.4/10
5.1/10
Exploit range
Attack complexity
Authentication
Adjacent network
Low
Single time
Confidentiality impact
Integrity impact
Availability impact
Partial
Partial
Partial
Affected software
Medtronic -> 2090 carelink programmer firmware 

 References:
https://ics-cert.us-cert.gov/advisories/ICSMA-18-058-01