Vulnerability CVE-2018-10634


Published: 2018-08-13

Description:
Medtronic MMT 508 MiniMed insulin pump, 522 / MMT - 722 Paradigm REAL-TIME, 523 / MMT - 723 Paradigm Revel, 523K / MMT - 723K Paradigm Revel, and 551 / MMT - 751 MiniMed 530G communications between the pump and wireless accessories are transmitted in cleartext. A sufficiently skilled attacker could capture these transmissions and extract sensitive information, such as device serial numbers.

Type:

CWE-319

(Cleartext Transmission of Sensitive Information)

CVSS2 => (AV:A/AC:M/Au:N/C:P/I:N/A:N)

CVSS Base Score
Impact Subscore
Exploitability Subscore
2.9/10
2.9/10
5.5/10
Exploit range
Attack complexity
Authentication
Adjacent network
Medium
No required
Confidentiality impact
Integrity impact
Availability impact
Partial
None
None
Affected software
Medtronic -> Minimed 530g mmt-551 firmware 
Medtronic -> Minimed 530g mmt-751 firmware 
Medtronic -> Minimed paradigm 508 insulin pump firmware 
Medtronic -> Minimed paradigm real-time mmt-522 firmware 
Medtronic -> Minimed paradigm real-time mmt-722 firmware 
Medtronic -> Minimed paradigm revel mmt-523 firmware 
Medtronic -> Minimed paradigm revel mmt-523k firmware 
Medtronic -> Minimed paradigm revel mmt-723 firmware 
Medtronic -> Minimed paradigm revel mmt-723k firmware 

 References:
http://www.securityfocus.com/bid/105044
https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-02

Copyright 2024, cxsecurity.com

 