Vulnerability CVE-2018-14786


Published: 2018-08-23

Description:
Becton, Dickinson and Company (BD) Alaris Plus medical syringe pumps (models Alaris GS, Alaris GH, Alaris CC, and Alaris TIVA) versions 2.3.6 and prior are affected by an improper authentication vulnerability where the software does not perform authentication for functionality that requires a provable user identity, where it may allow a remote attacker to gain unauthorized access to various Alaris Syringe pumps and impact the intended operation of the pump when it is connected to a terminal server via the serial port.

Type:

CWE-306

(Missing Authentication for Critical Function)

CVSS2 => (AV:N/AC:L/Au:N/C:P/I:P/A:P)

CVSS Base Score
Impact Subscore
Exploitability Subscore
7.5/10
6.4/10
10/10
Exploit range
Attack complexity
Authentication
Remote
Low
No required
Confidentiality impact
Integrity impact
Availability impact
Partial
Partial
Partial
Affected software
BD -> Alaris cc firmware 
BD -> Alaris gh firmware 
BD -> Alaris gs firmware 
BD -> Alaris tiva firmware 

 References:
http://www.bd.com/en-us/support/product-security-and-privacy/product-security-bulletin-for-various-alaris-plus-syringe-pumps-sold-and-in-use-outside-the-united-states
http://www.securityfocus.com/bid/105147
https://ics-cert.us-cert.gov/advisories/ICSMA-18-235-01

Copyright 2024, cxsecurity.com

 