Vulnerability CVE-2018-8861


Published: 2018-05-04

Description:
Vulnerabilities within the Philips Brilliance CT kiosk environment (Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior) could enable a limited-access kiosk user or an unauthorized attacker to break-out from the containment of the kiosk environment, attain elevated privileges from the underlying Windows OS, and access unauthorized resources from the operating system.

Type:

CWE-noinfo

CVSS2 => (AV:L/AC:L/Au:N/C:C/I:C/A:P)

CVSS Base Score
Impact Subscore
Exploitability Subscore
6.8/10
9.5/10
3.9/10
Exploit range
Attack complexity
Authentication
Local
Low
No required
Confidentiality impact
Integrity impact
Availability impact
Complete
Complete
Partial
Affected software
Philips -> brilliance ct big bore firmware 
Philips -> Brilliance firmware 64 
Philips -> Brilliance ict firmware 
Philips -> Brilliance ict sp firmware 

 References:
http://www.securityfocus.com/bid/104088
https://ics-cert.us-cert.gov/advisories/ICSMA-18-123-01
https://www.usa.philips.com/healthcare/about/customer-support/product-security

Copyright 2024, cxsecurity.com

 

Back to Top