Vulnerability CVE-2020-16224
Published: 2020-09-11
Description: |
Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, PerformanceBridge Focal Point Version A.01, IntelliVue patient monitors MX100, MX400-MX850, and MP2-MP90 Versions N and prior, IntelliVue X3 and X2 Versions N and prior. The software parses a formatted message or structure but does not handle or incorrectly handles a length field that is inconsistent with the actual length of the associated data, causing the application on the surveillance station to restart. |
Type:
CWE-130 (Improper Handling of Length Parameter Inconsistency )
CVSS2 => (AV:A/AC:L/Au:N/C:N/I:N/A:P)
CVSS Base Score |
Impact Subscore |
Exploitability Subscore |
3.3/10 |
2.9/10 |
6.5/10 |
Exploit range |
Attack complexity |
Authentication |
Adjacent network |
Low |
No required |
Confidentiality impact |
Integrity impact |
Availability impact |
None |
None |
Partial |
References: |
https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01
|
|
|
closedb();
?>
Copyright 2024, cxsecurity.com
|
|
|