Vulnerability CVE-2020-25156


Published: 2022-04-14

Description:
Active debug code in the B. Braun Melsungen AG SpaceCom Version L8/U61, and the Data module compactplus Versions A10 and A11 and earlier enables attackers in possession of cryptographic material to access the device as root.

Type:

CWE-489

(Leftover Debug Code)

CVSS2 => (AV:N/AC:L/Au:S/C:C/I:C/A:C)

CVSS Base Score
Impact Subscore
Exploitability Subscore
9/10
10/10
8/10
Exploit range
Attack complexity
Authentication
Remote
Low
Single time
Confidentiality impact
Integrity impact
Availability impact
Complete
Complete
Complete

 References:
https://www.cisa.gov/uscert/ics/advisories/icsma-20-296-02
https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html