Vulnerability CVE-2017-11578


Published: 2019-07-02

Description:
It was discovered as a part of the research on IoT devices in the most recent firmware for Blipcare device that the device allows to connect to web management interface on a non-SSL connection using plain text HTTP protocol. The user uses the web management interface of the device to provide the user's Wi-Fi credentials so that the device can connect to it and have Internet access. This device acts as a Wireless Blood pressure monitor and is used to measure blood pressure levels of a person. This allows an attacker who is connected to the Blipcare's device wireless network to easily sniff these values using a MITM attack.

Type:

CWE-200

(Information Exposure)

CVSS2 => (AV:N/AC:M/Au:N/C:P/I:N/A:N)

CVSS Base Score
Impact Subscore
Exploitability Subscore
4.3/10
2.9/10
8.6/10
Exploit range
Attack complexity
Authentication
Remote
Medium
No required
Confidentiality impact
Integrity impact
Availability impact
Partial
None
None
Affected software
Blipcare -> Wi-fi blood pressure monitor firmware 

 References:
http://packetstormsecurity.com/files/153225/Blipcare-Clear-Text-Communication-Memory-Corruption.html
https://github.com/ethanhunnt/IoT_vulnerabilities/blob/master/Blipcare_sec_issues.pdf
https://seclists.org/bugtraq/2019/Jun/8

Copyright 2024, cxsecurity.com

 