Vulnerability CVE-2017-12725


Published: 2018-02-15

Description:
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.

Type:

CWE-798

CVSS2 => (AV:N/AC:M/Au:N/C:P/I:P/A:P)

CVSS Base Score
Impact Subscore
Exploitability Subscore
6.8/10
6.4/10
8.6/10
Exploit range
Attack complexity
Authentication
Remote
Medium
No required
Confidentiality impact
Integrity impact
Availability impact
Partial
Partial
Partial
Affected software
Smiths-medical -> Medfusion 4000 wireless syringe infusion pump 

 References:
http://www.securityfocus.com/bid/100665
https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A

Copyright 2024, cxsecurity.com

 